3282 Power Factor RLC Load

PRODUCT


 
MODEL 3282
Power 1875W / 3700VA
Reactive Power
(With Extend OPTION)
C : 900 VARmax @ 110V / 60Hz    L : 400 (1200) VARmax @ 110V / 60Hz
C : 3785 VARmax @ 220V / 60Hz  L : 800 (3200) VARmax @ 220V / 60Hz
Current (Ampere) 18.75 Arms / 46.875 Apeak
Reactive Current
(With Extend OPTION)
C : 8.6 Amax @ 110V / 60Hz        L : 3.6 (10.8) Amax @ 110V / 60Hz
C : 17.2 Amax @ 220V / 60Hz      L : 3.6 (14.4) Amax @ 220V / 60Hz 
Voltage (Volt) 50~280 Vrms / 400Vdc
 
특징
 
4 digit V / A/W Meterdisplay the VoltageVrms, Vpeak, Vmax., Vmin)、Current Irms, Ipeak, Imax., Imin.)、Watt,  Voltampere VA)、FrequencyCrest FactorPower FactorTotal Harmonic Distortion of VoltageVTHD, Voltage HarmonicVH)、Total Harmonic Distortion of CurrentITHD, Current HarmonicIH
CC, Linear CC, CR, CV, CP and AC Rectifier Load mode
Crest factor range : 1.414~5.0
Power factorPF) range : 0~1 lead or-1~0lag
Inductive / capacitive load power factor range: 0 ~ 1 Current waveform leads or lags against voltage waveform
Built-in function test modes include UPS Efficiency, PV Inverter Efficiency, UPS Back-up time, Battery Discharge time, UPS transfer time, Fuse/Breaker Trip/Non-Trip, Short circuit , OCP, OPP test modes
Turbo mode is able to increase to 2 times the current and power of electronic load in a short period which is the most suitable for Fuse / Breaker test and short circuit, OCP, OPP test of AC power supply
Time measurement can be applied to batteries, UPS, fuses and circuit breakers and other tests
Eight units parallel up to 15KW and three-phase or Y load connection can be synchronized control by onemaster unit
For inductive or capacitive loads of general household and industrial electrical equipment, various electrical equipment that require load currents lagging or leading voltage, such as electric fans with motors, air conditioners with compressors, refrigerators, etc., 3282 RLC power factor AC load provided the power factor load of household and industrial electrical appliances can be simulated, and the anti-islanding test is provided for the grid-connected function of solar inverters, electric vehicle inverters and energy storage devices. The required real power and virtual inductive power and capacitve power can be set independently. 3282 single 280V/18.75A capacity can reach 1.875KW/3200VA max, as long as the load parameters of various electrical equipment and inverters are input into 3282, the inverter/UPS AC power output to be verified is connected to 3282 The AC load can simulate various electrical equipment with inductive/capacitive power factor loads.
Support on-load boot; at first set Load ON to support on-load boot, inverter or uninterruptible power supply is turned on directly with the set load current, used to verify whether the starter is stable when the Inverter is connected.
Supports the loading and unloading angle control; the loading and unloading angle control, the full range of 0-359 degrees can be set to verify whether the Inverter output voltage transient response is stable when the actual electrical plugging and unplugging, and whether Overshoot/Undershoot is within the allowable range.
Support positive half-cycle or negative half-cycle loading; used to verify whether the Inverter output voltage remains stable when the actual appliance has only positive half-cycle or negative half-cycle load current.
Supports SCR/TRIAC current phase modulation waveforms, 90 degree Trailing edge and Leading Edge.
Supports the Inrush Current of the inverter at startup and the Surge Current test when the load is suddenly plugged in (Hot Plug-in) during testing.
Frequency Range : DC, 40~70Hz
Voltage and current monitoring
Can be controlled by external voltage for CC, Linear CC, CR, CV, CP operating modes
Protection against V, I, W, and
Optional interfaceGPIBRS232USBLAN
The most complete measurement capabilities 3282 power Factor RLC Load built-in 16-bit A/D and DSP precision measurement circuit, provides accurate measurements, measurement items have Vrms, Arms, Watt, VA, CF, PF, THD, VTHD, ITHD, Ipeak, Amax, Amin, Vmax, and Vmin In addition to these measurement functions, it also provides time measurementproducts such as UPS, fuses and circuit breakers etc. trip or blow time and transfer time for Off-line UPS